DOOR REGULATION
Description
If the door is not closed and Error 06 appears on the screen during the sterilization cycle, check and regulate the door with its door key. Follow the steps below and watch the video-tutorial to learn how
Activities: | 6 |
Duration: | 11 min |
Difficulty: | Low |
Necessary Tools
- Door key
Replacement Parts
- No parts required
DOOR REGULATION
Step 1
Open the door

Please open the machine door
Step 2
Regulation of the door with the door key

Please insert the key in the door slot
Step 3
Regulation of the door with the door key

Please hook the key on the rotation nut
Step 4
Regulation of the door with the door key

Please turn the key in the counterclockwise direction to move the door plate closer to chamber
Step 5
Close the door

Close the door
Step 6
Check the technical action done

Check the technical action done